Optical Bubble Microflow Meter for Continuous Measurements in a Closed System
Oglądaj/ Otwórz
Data
2024Autor
Rosiak, Michał
Stanisławski, Bartłomiej
Kaczmarek, Mariusz
Metadata
Pokaż pełny rekordStreszczenie
This paper describes the design, operation and test results of a simple microprocessor-based device for measuring slow liquid flows. The device uses a module of 30 digital optical sensors to track the movement of a single air bubble inserted into a tube of flowing liquid. During a measurement session, the air bubble remains within the sensor module at all times, allowing the instrument to take measurements for any length of time. The liquid whose flow rate is being measured moves only in the closed tube system, without contact with other components of the device. The test of the device itself was carried out using a tube with an inner diameter of less than 1 mm, where the device is capable of measuring flow rates on the order of microliters per minute. Tests of the device showed good agreement between the measured volumetric flow rate and the reference flow rates of the infusion pump over the entire measurement range. The advantages and limitations of the device are discussed, as well as the prospects for developing the method.
Collections
- Artykuły naukowe [47]
Z tą pozycją powiązane są następujące pliki licencyjne: